NR7200 series SF6 gas micro-water tester

Used to measure trace water content in SF6 gas, it features high measurement accuracy, wide range, and simple testing. The instrument introduces a high-precision temperature sensor, with high accuracy, good repeatability, fast testing speed, and small drift. The test results can be automatically converted to standard values at 20℃. It has automatic zero calibration and full-range curve tracking correction functions, a pioneering curve display, power display, and advanced probe protection function, making testing simpler.


Messages

Product patent

Product description

Product Introduction
  Used to measure the trace water content in SF6 gas, it has the advantages of high measurement accuracy, wide range, and simple testing. This instrument introduces a high-precision temperature sensor, which has the advantages of high accuracy, good repeatability, fast testing speed, and small drift. The test results can be automatically converted to standard values at 20℃. It has automatic zero calibration and full-range curve tracking correction functions, a pioneering curve display, power display, and advanced probe protection functions, making testing simpler.
 
Main Features
  ◆ High-precision humidity sensor
  The instrument uses a high-precision humidity sensor, which has the advantages of high accuracy, good repeatability, fast testing speed, and small drift, making it an ideal choice for industrial environments with low dew points and the need to control dry points;
  ◆ Anti-interference
  It adopts a dual-power channel design, separating the electrolytic electrode and the measuring electrode, automatically removing various interferences, and greatly improving the precision of the test results. The measuring electrode data is directly converted by A/D sampling, and the true value of the measured data eliminates the errors caused in the analog circuit, making the test results more accurate;
  ◆ Powerful Functions
  Automatic zero calibration and full-range curve tracking correction, the test results can be automatically converted to standard values at 20℃. It has a chemical cleaning option, which allows the product to perform accurate and stable measurements in environments with high concentrations of chemicals and cleaning agents, thus ensuring accurate measurements between each calibration interval.
 
Technical Parameters
Measurement Range
-80℃ ~ +20℃
Accuracy
≦1℃
Resolution
0.01℃ or 1PPm
Output Interface
RS232
Response Time
When the dew point is +60℃ ~ -20℃, ≦1min
-20℃~ -40℃, ≦3min
-40℃~ -60℃, ≦5min
Operating Environment
Temperature: -20℃ ~ +60℃
Relative Humidity: 0 ~ 95%
Pressure: ≤1.0Mpa

 